Worldwide provider of
test and leak detection
equipment to manufactures
since 1981.

Contact Us >>

Home » Products » Leak Test Instrumentation » Sentinel I28 » Sentinelâ„¢ I28
Sentinel I28 Leak Test Device
Click to enlarge

Quote Request Options:

Test Type: 
Pressure Selections: 
Mass Flow Range: 
Reject Rate: 
Reject Units: 
Pressure: 
Pressure Units: 
Quantity:
   

Related Files:
Download PDF of this page Spec Sheet
Download PDF of this page I28 Brochure
Download PDF of this page 2D Drawing
Download PDF of this page 3D Drawing

Sentinelâ„¢ I28

The Sentinel I28 is the world’s most advanced multi-functional test instrument on the market. The I28 features a standard valve manifold design tested with over 40,000,000 cycles. CTS pairs this manifold with a new high performance pressuretransducer that reduces test cycle and improves accuracy and repeatability.

Advanced Product Features

  • Wall mount configuration 12” W x 9.25” H x 8.75” D
  • Powerful 32-bit processor & 24 –bit A/D converter
  • Language neutral color LCD 12 line 480 x 272 operator display
  • XY Graphing (pressure, loss, or flow)-English, Chinese, Korean, Spanish
  • Indicator Light bar identifies test status Pass/Fail/In-test
  • Barcode ready input RS232 Alpha-numeric string stored with test result. 
  • Modular pneumatic and control packages
  • Single test channel

Multi-test Configurable for Pressure Decay or Mass Flow Test Requirements                

  • Pressure/Vacuum decay-pressure change
  • Pressure/Vacuum decay-with certified mass flow leak standard
  • Pressure/Vacuum decay-pressure change/time
  • Occlusion test (vacuum or pressure test)
  • Mass flow direct flow reading (pressure test only)
  • Mass flow continuous direct flow reading (pressure test only)
  • Mass flow with certified mass flow leak standard (pressure test only)
  • Volumetric Test (pressure test only)-“under development
  • “Parent Program” holds a customized test sequence for your multi-test application
  • Storage for 99 pre-programmed test sequences
  • Test Program can be selected through keypad, RS232, or Binary
  •  
  • Test results evaluation is Selectable (Fail/Pass/Fail etc)
  • Test result output format is configurable format

Selectable Test Parameters

  •  Quick Test
  •  Test Pressure Level Compensation (selectable on/off)
  • Adjustable Environmental Drift Correction per test program

Standard Test Pressure Ranges

      Standard pressure transducer, valve manifolds, and regulator ranges:

          -14.5, -2, -0.1 to 4”HG, 0.5, 2, 30, 100, 200, 400, 500 or electronic regulation (PSIG)

  • Optional low volume test manifold available
  • Optional High cV flow valves and manifold available up to 100 psig range allows 5 times faster fill and exhaust to reduce cycle times on large parts
  • Optional Low volume (0.07cc) manifold available
  • Absolute pressure transducer with accuracy of +/-0.05% of full scale
  • Six point pressure calibration
  • Data resolution 0.00001 psi/ and scc/m.
  • Optional electronic regulator available with multi-point calibration

Mass Flow Sensor Ranges

 CTS will optimize the flow sensor & flow range per application. Contact Application Engineering for available mass flow sensors.

Selectable Test Units of Measure

  • Pressure

PSIG, PSIV, inH2O, cmH2O, mmH2O, kg/cm2, ATM, inHg, cmHg, mmHg, kPa,  Pa, MPa, Bar,  mBar, Torr

  • Flow

Sccm, sccs,  scch,  slpm,  slps,  slph, scfm, scfs, scfh

  • Time

Sec, msec, min

  • Units of measure can be instrument global or per test program

Auto-Setup, Automatic Calibration, and Self Test

  • Auto setup, verifies test pressure and selects the best test timers for optimum results in the allowed time cycle (US Patent #7,594,424 B2)
  • Auto Calibration includes mass flow leak standard certified for pressure and flow
  • Transducer zero and span are automatically reset during auto calibration
  • “Performance factor” calibration feedback ensures accurate results
  • Manual edit of Calibration Program permitted
  • Self Test isolates the test instrument pneumatics to verify a “no leak” condition
  • Features ensure accurate results and easy operator setup and troubleshooting
  • Firmware validation document included

Tooling Control

  • 12 inputs/12 outputs with up to five (5) programmable tooling motions with feedback
  • Part marking control
  • Tooling I/O is  externally connected
  • I/O 24 VDC – optional I/O cable “Cable,CTS, A12-610(standardx2)”

Test Data Storage

  • Up to 5000 LIFO part results including Date / Time / Barcode / Test Results
  • Selectable part serialization in the test data stream
  • SPC data tracking for part traceability
  • Expandable download through USB port

Electrical Requirements and Connectivity

  • 88-264 VAC, 47-63 Hz or optional 22-26 VDC
  • RS232(2)
  • Telnet(1)
  • Email Alerts
  • Barcode ready (RS232)
  • Mobile text
  • USB mass storage device for backup, restore, and reports (FAT32, USB1.0,1.1,2.0)
  • ¼” Colder fitting for test line attachment

All CTS instruments include password protected program and test result storage. All CTS instruments are CE certified and carry a two year limited warranty.  Normal delivery for all standard instruments is 2-4 weeks plus delivery time. Contact Us to select the proper standard test instrument for your application.

CTS has more than 30 years of experience perfecting high precision, repeatable and reliable instruments to optimize the most demanding leak test applications in any manufacturing environment. CTS application engineers can help you develop a custom solution specifically suited to your needs. See the “Custom Instrument” section for a brief overview of our capabilities.

You can trust CTS to provide you with the right solution to fit your specific testing requirement.

 
 
© 2014 Cincinnati Test Systems    •    Corporate Headquarters    •    10100 Progress Way    •    Harrison, OH 45030
Phone: (513) 367-6699
Privacy Policy         Site Map
Site Credits: Ecreativeworks